Horiba Medical
ABX Pentra 60 C+ Overview
5-Part Differential analysis Closed tube sampling…Workstation included
- Data management on stand-alone PC.
- Micro-sampling from whole blood (CBC: 30 µL – DIFF: 53 µL).
- Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) – DHSS* Technology.
- Perfect homogenization of blood samples with reagents – MDSS** Technology.
Horiba Medical Pentra 60 C+ Request For Quote
Features
Concept and Technology
- MDSS*
- DHSS*
- No compressor
- No shear valve, less maintenance
- Microsampling of 30 μL (CBC) or 53 μL (CBC+DIFF)(Exceptional results with all sample types, even very small volumes Pediatric, Oncology, etc.)
- Data management on external PC Stand alone capability
- Windows Platform Easy to use
- Closed tube sampling Reduces biohazard risk
- External barcode reader 100% secure sample identification
- 26 parameters
- WBC, RBC and Platelet histograms
- Color matrix for WBC Differential.
- Pathological and morphological alarms
- WBC Differential performed by DHSS technology
- WBC, RBC and Platelet histograms
- Basophil measurement performed in a specific channel
- Percentage and absolute value of neutrophils, eosinophils, basophils, lymphocytes and monocytes
- Determination of 2 additional sub-populations (% and #): Atypical lymphocytes (ALY)*, Large immature cells (LIC)*. * For Laboratory Use only (not FDA approved)
MDSS Microsampling
- Sampling : only 30 μL whole blood for CBC (53 μL for CBC + DIFF)
- Sample dispensed into pre-heated analysis chamber for highly reproducible results
- Tangential flow reagent dilution for optimal sample mixing
- * ABX Patents
Cytochemistry
- Sample incubation in a temperature-controlled chamber and enzymatic staining with Chlorazol Black. This reagent specifically stains leukocyte nuclei, granules and membranes.
Cytometry
- Injection of the prepared sample into a double hydrofocus cytometer (ABX patent) and analysis of cell complexity with a polychromatic light source
- Measurement of cell volume by impedance
- Measurement of cell content by cytochemistry and optical absorbance
Specifications
- DIMENSIONS & WEIGHT: Height Wiidth Depth Weight Analyzer 82 cm 57 cm 54cm 55 Kg 33,4 in 23,3 in 22 in 122 lb
- PRINTER: Laser
- PTHROUGHPUT: Up to 80 samples/hour in stat mode or in automatic mode
- SOUND PRESSURE LEVEL: < 60 dBa
- OPERATING TEMPERATURE & HUMIDITY: 16 – 34°C (61 – 93°F) room temperature Maximum relative humidity 80% for temperature up to 31°C (88°F) linearly decreasing humidity at 40°C (104°F)
- SPECIMEN VOLUME: CBC 30 μL / CBC + DIFF 53 μL
- POWER REQUIREMENTS: Power supply from 100 V to 240 V (± 10 %) 50 Hz to 60 Hz, Power consumption Maximum 230 W
- • REAGENTS: 5 reagents only: ABX Diluent, ABX Alphalyse or cyanide-free lyse (optional), ABX Cleaner, ABX Eosinofix, ABX Basolyse II
- MULTI DISTRIBUTION SAMPLING SYSTEM “MDSS”
- RBC/PLT DETECTION PRINCIPLES Method Impedance, Aperture diameter 50 μm, Counting depression 200 mb, Counting duration 2 x 5 seconds, Dilution ratio 1/10 000, Reaction temperature 35°C
- HGB MEASUREMENT Method Photometry, Wavelength 555 nm, Dilution ratio 1/250, Reaction temperature 35°C
- HCT MEASUREMENT Method Numeric integration
- WBC & BASO COUNT Method Impedance, Aperture diameter 80 μm, Counting depression 200 mb, Counting duration 2 x 6 seconds, Dilution ratio 1/200, Reaction temperature 35°C
- LEUKOCYTE DIFFERENTIATION Method Focused-flow impedance, Cytochemistry, Aperture diameter 60 μm, Diameter of the flow 42 μm, Injection duration 12 seconds, Dilution ratio 1/80, Incubation duration 12 seconds, Reaction temperature 35°C
- MCV, MCH, MCHC, RDW, PCT*, PDW* Calculation parameters
- DATA PROCESSING: 15” Flat Screen LCD, Motherboard: Intel Processor w/ SIS Chipset, On Board audio & video, Capacity: 10,000 results + graphics, OS: Windows NT 4.0 w/Service Pack 6A, PC: Intel Celeron 1.8 GHz (min.), RAM: 256M min., Hard Disk (40 Gb) min., Floppy Disk and CD ROM Drive, RS232C, User defined flagging Limits, Transmit patient and QC to LIS connection, Uni & Bi-directional connections, ABX and ASTM interfacing protocol, HemaLink Data Management readys
- QUALITY CONTROL MANAGEMENT: 12 selectable QC files, XB: 60 operator selectable files with statistics (20 samples per file) Within run, Levey-Jennings graphs, Unlimited QC results storage with HemaLink™
- LOGS: Reagents, calibration, maintenance, errors, blank cycle
- 26 PARAMETERS: WBC RBC PLT N# & N% HGB MPV L# & L% HCT PCT* M# & M% MCV PDW* EOS# & EOS% MCH BAS# & BAS% MCHC ALY*# & ALY*% RDW LIC*# & LIC*%
- LINEARITY: (VERSION V2.3) Parameters Linearity range Visible range** Units WBC 0 – 120 120 -150 103/μL RBC 0 – 8 8 – 18 103/μL HGB 0 – 24 24 – 30 G/dL HCT 0 – 67 67 – 80 % PLT (whole blood) 0 – 1900 1900 – 2800 103/μL PLT (plt concentrate mode) 0 – 2800 2800 – 3200 103/μL
- PRECISION: Parameters %CV Range WBC < 1.5 4.0 – 11.0 x 103/μL RBC < 1.5 4.0 – 6.0 x 106/μL HGB < 1.0 11.0 – 18.0 g/dL HCT < 1.5 35 – 55 % RDW < 2.0 80 – 100 PLT < 5.0 150 – 400 x 103/μL MPV < 3.0 7.6 – 10.9
- ACCURACY: Mean % Mean Parameters Difference Difference WBC < 3 ± 0.2 RBC < 3 ± 0.10 HGB < 3 ± 0.3 HCT < 4 ± 1.5 PLT < 5 ± 10