micros_60_usa

Horiba Medical

ABX Micros 60 Overview

  • Impedance technology, photometry
  • Microsampling: Only 10 µL
  • Smart card based memory storage (optional)
  • No compressor (stepper motor technology)

Horiba Medical Micros 60 Series Request For Quote

  • This field is for validation purposes and should be left unchanged.

Features

pdf

Concept and Technology
    • CBC & 3-DIFF: Impedance Technology & specific Lyse action
    • ABX Liquid valves: Precise volumes and reliability, without pinch valves
    • No compressor: Stepper motor, dependable and low noise
    • Easy to Operate: Single button startup/shutdown
Quality Control Features
    • Load assay values from disc
    • Westguard Rules:• Real Time (1-2S & 1-3S)• On Demand (2-2S, R-4S, 4-1S, 10x)
    • Selectable formats for viewing QC performance
Review and Edit
    • Comprehensive sample tracking and routing
    • Abnormal results are highlighted
    • Sample rerun capability with display of multiple runs on single screen
    • Customizable offline testing fields
Flexible Patient Reporting
    • Customizable cumulative or individual reports available
    • Output to printer and /or archived .pdf-file
Delta Checking
    • Color flags, execution of rules based upon failure
Increased Efficiency
    • Virtually limitless storage of patient data
Remote QC capability
  • Simple remote workstation configuration
  • Auto-connect for QC viewing and printing

Specifications

    • Analyzer Dimensions Height Width depth Weight 16.5” 14.2” 12.6” 30.9 lb
    • Management System 2.75 18.25” 8.5” 43.0 lb
    • Printer: Digital Laser
    • Throughput: Up to 55 samples/hour in Closed Tube • Up to 60 samples/hour in Open Tube (sold separately)
    • Sound Pressure Level: <60 dBa
    • Operating Temperature & Humidity: 16 – 34°C (61 – 93°F) room temperature Maximum relative humidity 80% for temperature up to 31°C (88°F) decreasing linearity humidity at 40°C (104°F)
    • Specimen Volume: 10 μL
    • Power Requirements: Power supply from 90 Vac to 240 Vac 50 Hz to 60 Hz Power consumption Maximum 230 VA
    • Reagents: – ABX Minidil LMG – ABX Alphalyse-360 – ABX Miniclean – ABX Minoclair
Methods & Technologies
    • RBC/PLT detection principles Method Impedance • Ruby diameter 50 μm • Depression of count 200 mb • Duration of count 2 (or 3) x 6 seconds • Dilution ratio 1/15 000
    • HGB measurement • Method Photometry • Wavelength 555 nm • Dilution ratio 1/250
    • HCT measurement Method Numeric integration
    • WBC measurement • Method Impedance • Ruby diameter 80 μm • Depression of count 200 mb • Duration of count 2 (or 3) x 6 seconds • Dilution ratio 1/250
    • 3 part Differentiation • Method Impedance + specific lyse action
    • MCV,MCH,MCHC,RDW,PCT*,PDW* • Calculation
Certification UL CSA
    • NF EN 60601-1. UL 3101-1 C22.2 n°1010-1 / NF EN 61326. B / NF EN 61000-3-2 / NF EN 61000-3-3
System Specifications
    • DELL OptiPlex Small Form computer 53 GHz, 3M 1 GB RAM, 80 GB Hard Drive, 8X CDRS-DVD 3.5″, 1.44MB Floppy disk Drive
    • 17″ flat-screen monitor
    • Microsoft WINDOWS XP PRO
    • PC AnyWhere
    • Adobe PDF Creator
Parameters & Performance Data
    • •16 Parameters,WBC RBC PLT, LYM# & LYM% HGB MPV, MON# & MON% HCT, GRA # & GRA% MCV, MCH, MCHC, RDW
    • Linearity: Parameter Linearity Limit Visible Range, WBC 0 – 100 x103/mm3 100 – 150 x103/mm3, RBC 0 – 8 x106/mm3 8 – 18 x106/mm3, HGB 0 – 26 g/dL 26 – 30 g/dL, HCT 0 – 80 % 80 – 90%, PLT (Whole blood) 0 – 2200 x103/mm3 2200 – 6000 x103/mm3, PLT (PLT concentrate) 0 – 4000 x103/mm3 4000 – 6000 x103/mm3
    • •Precision: Parameter %CV at, WBC <2.5 10 x103/mm3, RBC <2 5 x106/mm3, HGB <1.5 15 g/dl, HCT <2.0 45%, PLT <5.0 300 x103/mm3, LYM% <5.0 40%, MON% <10 10%, GRA% <3 50%
    • •Carry-over: WBC <0.5% 63.0 x 103/mm3, , RBC <0.5% 7.58 x 103/mm3, HGB <0.5% 23.4 g/dL, PLT <0.5% 988 x 103/mm3
Measurement Principle
  • Bi-chromatic light absorbance
  • Choice of 8 wavelengths from 340 to 700 nm
  • Wavelength selection with filter wheel
  • Sensor
  • Photo diode