DxH 690T
Hematology Analyzer
Trusted Quality for Mid-Volume Labs
The DxH 690T is a bench-top hematology analyzer that delivers trusted results for mid-volume labs. With this table-top configuration, you can count on best-in-class hematology technology, minimal slide reviews, maximum efficiency, and significant cost savings.
This product may not be available in your country or region at this time. Please contact your Beckman Coulter sales representative or distributor for more information.
Quality That Counts
Trust your results—the first time—with a high-quality benchtop hematology analyzer that delivers precise, accurate results and advanced cellular insights.
- Achieve superb RBC, PLT, and WBC differentials through near-native-state cellular characterization
- Deliver high-confidence results on the first run—minimizing reruns or reflex tests
- Improve slide interpretations with intelligent technology that helps optimize smearing, ensuring consistent slide quality
Productivity That Liberates
Free up your mid-size laboratory’s hematology team and empower your busy staff to focus on samples and tasks that truly need attention.
- Simplify workload planning with predictable hematology workflows that let you save time for the unexpected
- Optimize your limited resources with reliable flagging
Savings That Empower
Tap into powerful savings to help your lab devote time, budget, and resources where they are needed most.
- Streamline QA/QC processes through customizable, automated technology with daily reminders and remote support
- Save time, reduce costs, and optimize hematology instrument usage with fewer reflex tests
- Reduce hematology reagent use and associated costs
- Reallocate up to 200 minutes of daily run-time previously spent on reflex testing1
- Reduce overhead costs like electricity and maintenance
SPECIFICATIONS
Methodology |
Enhanced Coulter Principle, VCS 360 with DataFusion | |||||
ParametersEmpty heading
|
CBC: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Differential: NE%, LY%, MO%, EO%, BA%, NRBC%, NE#, LY#, MO#, EO#, BA#, NRBC# Retic: RET, RET#, MRV, IRF Body Fluids (Cerebrospinal, Serous, or Synovial): RBC, TNC |
|||||
Throughput (tests per hour) |
Up to 100 samples/hour | |||||
Sample Volumes |
Whole Blood: 165 μL aspiration volume. Predilute: 50 μL whole blood into 200 μL of diluent | |||||
Analytical Measuring Ranges1 |
Whole Blood (CBC) | |||||
| Parameters | Units | Analytical Measuring Range | ||||
| WBC | x103 cells/μL | 0.050–400.000 | ||||
| RBC | x106 cells/μL | 0.005–8.500 | ||||
| HGB | g/dL | 0.10–25.50 | ||||
| MCV | fL | 50.00–150.00 | ||||
| RDW | % | 10.00–40.00 | ||||
| RDW-SD | fL | 15.00–150.00 | ||||
| PLT | x103 cells/μL | 3.0–3000.0 | ||||
| MPV | fL | 5.00–25.00 | ||||
| Body Fluids | ||||||
| TNC | cells/mm3 | 20–89,000 | ||||
| RBC | cells/mm3 | 1,000–6,200,000 | ||||
Data Management |
Up to 50,000 patient results with histograms, scatter plots, and demographics
100,000 patient results with graphics for power computer |
|||||
Quality Assurance |
QC with Levy-Jennings graph; XB/XM for moving averages, daily check, intelligent quality monitoring, MRV, IRF, customizable calibration and QC reminders and alerts, auto export of QC | |||||
Acoustic Noise Level |
≤ 60 dBa | |||||
Radio Frequency and Interference Immunity |
Meet the product level standards EN 61326-1 (Electrical Equipment for Measurement, Control, and Laboratory Use—EMC Requirements), and EN 61326-2-6 (Part 2, Specific Requirements for In-Vitro Diagnostic Devices) | |||||
Power Consumption |
DxH 690T SPM: 520 W (1775 BTU/hour), Monitor: 35 W (120 BTU/hour), DxH Power Computer: 160 W (546 BTU/hour) | |||||
Dimensions and weight2 |
Width | Height | Depth | Weight | ||
| 76.2 cm (30 in.) | 90.17 cm (35.5 in.), with cover closed
113.03 cm (44.5 in.), with cover open |
83.82 cm (33 in.), including the removable back panel
80 cm (31.5 in.), excluding the removable back panel |
~141.97 kg (313 lbs.), including monitor ~136.98 kg (302 lbs.), |
|||
Additional Clearance |
15.2 cm (6.0 in.) per side of instrument for troubleshooting; 3.8 cm (1.5 in.) behind instrument for sufficient cooling | |||||
Maintenance and Troubleshooting
Service |
New Preventative Maintenance (PM) schedule with customer care visits |
Support |
Onboard IFU and Instructional Videos, Automated QC and calibration reminders, PROService |
Analyzer Downtime |
Minimum of 30 minutes in cleaner every 24 hours with automatic daily checks |
Research Use Only Parameters
Available RUO Parameters3Empty heading |
BFM, BFM#, BFP#, EGC, EGC#, HLR, HLR#, LHD, MAF, MSCV, PCT, PDW, RDWR, RDWR-SD, RSF, UGC, UGC#, UWROP, WDOP, WNOP, WROP, CPD (refer to reference 3 for additional RUO parameters) |
Research Use Only. Not for use in In Vitro Diagnostic procedures.
1. DxH 690T IFU (PN C06947)
2. DxH 690T ReadMe (PN C53971)
3. DxH 690T RUO Addendum
Tell Us What Your Looking For?
"*" indicates required fields
Looking For Reagents?
Explore our range of high-quality reagents and consumables for accurate laboratory testing.